New TopShim run:

version = topshim 2.4.0
date = 2024-05-16
time = 12:35:27.326
command = topshim 

FINE SHIMMING

Parameters:

probe = Z113652_0298
solvent = DMSO
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 2.49 ppm

Results:


1D Shimming:
Iteration  1: FWHM Before= 6.7 Hz, Target= 0.5 Hz, After= 0.9 Hz
Iteration  2: FWHM Before= 0.9 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  3: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  4: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  5: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  6: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  7: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.5 Hz
Iteration  8: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 0.3 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.485

shim changes: 
Z	-157
Z2	-113
Z3	+174
Z4	+503
Z5	-661
Z6	-1229
Z7	+6966
Z8	-3604

duration = 1 min 53 sec
completed successfully
finished Thu May 16 12:37:19 2024


========================================================================